Workstation vertical

Arbeitsstation

System details

  • Front sash with individual settings
  • Light curtain for activity monitoring
  • Integrated lighting
  • Monitoring of filter pressure and differential pressure
  • Integrated pass-through
  • Monitoring of temperature
  • Programmable logic controller (PLC) for monitoring and controlling function, ventilation, equipment
  • Filters for supply air according ISO class

  • Connections for hotplates
  • Bottling station (various versions)
  • Underfloor flushing
  • Dropping
  • Media supply acc. customer requirements
  • Remote maintenance
  • Control units for media (hand, foot, display)
  • Integrated ultrapure water system (various manufacturers)
  • Still
  • Ultrasonic basin
  • Base cabinet with revolving doors or drawers
  • Exhaust air scrubber and exhaust air irrigation

Laminar flow

DIN EN 14175

Modular design

Reduce energy consumption

MK Versuchsanlagen workstation vertical

Our metal-free systems are characterized by their modular design, enabling us to create custom-made solutions tailored to your specific needs:

Top Module: Regulated ventilation system with filtered laminar airflow
Middle Module: Customizable work surface and front sliding panel
Base Module: Flexible substructure for additional storage space or legroom

The metal-free construction allows for contamination-free sample preparation at the highest level, for example, in trace analysis applications.

Product features

  • Supply air via available supply air from in-house ventilation system
  • Supply air is fed into the working area via terminal filter
  • Vertical laminar flow is realized by a fine-meshed monofilament fabric
  • Monofilament fabric can be changed
  • Laminar air flow in the whole working area
  • Control of supply air, recirculated air and exhaust air via the PLC ensures the correct pressure in the working chamber
  • Compliance with DIN EN 14175
  • Exhaust air can be discharged via air scrubbers or directly transferred to the exhaust air system (depending
    on the used chemicals)
  • Room air can also be extracted via the base module and fed into the ventilation system (‘recirculating air’)
Technical data
Materials PP, PET, POM, PFA, PVDF
Operating voltage 400 V, 5-pin
Air volume requirements Supply air min. 400 m³/h, exhaust air min. 450 m³/h (Width: 1200 mm)
Air velocity in working area

ca. 0.15 m/s

Terminal supply air filters

High-performance HEPA fi lter, fi lter class H13 to U16 (depending on requirements) with pre-filter F9

Cleanroom classes according DIN EN ISO 14644-1

Class 3 or better (0.2 μm, 0.3 μm) „at rest“

Compliance with the DIN EN ISO 14175

Defined safety specifications of the retention and air exchange capacity for personal and product protection

Illuminance

> 750 Lux

Interfaces

Ethernet or other common fieldbus systems and serial interfaces

Dimensions

1200 x 975 x 2260 mm (W x D x H), Widths 900 mm, 1500 mm and 1800 mm also available

Weight (without equipment)

260 kg for 1200 mm width (320 kg, 380 kg for higher widths)

Dimensions working area

1180 x 690 x 880 mm (W x D x H), width varies with total width

Working height

700 mm, 900 mm

Front sash opening height

Individually adjustable via PLC

Contact

Contact us for guidance on your project

Talk to our experts.

More products