Combination Station vertical

System details

  • Automatic vertical front sash (clear)
  • Light curtain for activity monitoring
  • Integrated LED lights
  • Monitoring of differential pressure and filter loading
  • Evaporation chamber with integrated pass-through
  • Temperature monitoring of exhaust air
  • Monitoring of temperature
  • Programmable logic controller (PLC) for monitoring and controlling of operation, airflows and equipment
  • Supply air filters according to ISO class
  • Volume flow controllers for supply and exhaust air

  • Connections for hotplates
  • Bottling station (various versions)
  • Exhaust tubflushing
  • Drain / wastewater connection
  • Customized media supply (interior and exterior)
  • Remote monitoring and servicing
  • Integrated ultrapure water system (various manufacturers)
  • Integrated acid stills
  • Ultrasonic basin
  • Base cabinet (vented) with hinged doors or drawers
  • Exhaust air scrubber and exhaust air irrigation

Laminar flow

DIN EN 14175

Modular design

Reduce energy consumption

MK Versuchsanlagen combination Station vertical

Our metal-free systems feature a modular design, allowing us to create custom-made solutions tailored to your specific needs:

  • Top module: Controlled ventilation system with filtered laminar airflow

  • Middle module: Customizable work surface and front sash

  • Base module: Flexible substructure for additional storage space or legroom

The metal-free construction enables contamination-free sample preparation at the highest level, ideal for trace analysis, for example. The station consists of two self-contained work areas with laminar airflow: the work chamber and the evaporation chamber, which are connected by a pass-through. The two autonomous chambers allow the evaporation processes to be separated from subsequent steps, ensuring improved personnel protection.

Additional features

  • Metal-free construction (no exposed metal) to prevent contamination and corrosion

  • Supply air via available supply air from in-house ventilation system

  • Vertical laminar flow achieved though a fine-meshed monofilament fabric diffuser

  • Laminar airflow throughout the entire working chamber; air is exhausted through perforated work surface

  • Control of supply air and exhaust air via the PLC ensures the correct pressure in the working chamber

  • Exhaust air is extracted through a perforated fl oor via underfloor basin

  • Compliant with DIN EN 14175-3 safety and performance standards (containment, air exchange effi ciency)
  • Simultaneous protection of personnel and samples/products
  • Outstanding air and energy effi ciency due to optimized and smart airfl ow management
  • Exhaust air can be discharged via air scrubbers or directly transferred to the exhaust air system
  • Room air can also be extracted via the base module and fed into the ventilation system (‘recirculating air’)
  • Separate chamber (with independent vertical laminar flow) for isolated evaporation / hotplate usage
Technical data
Materials PP, PET, POM, PFA, PVDF
Operating voltage 400 V, 5-pole
Air volume requirements

Supply air min. 400 m³/h, exhaust air min. 450 m³/h (Width: 1200 mm)

Flow velocity in working area approx. 0.15 m/s
Terminal supply air filters High-performance HEPA/ULPA filter, class H13 to U16 (depending on requirements)
Cleanroom class according
DIN EN ISO 14644-1
Class 3 or better (0.2 μm, 0.3 μm, 0.5 μm, 1 μm, 5 μm) „at rest“
Illuminance > 750 Lux in both chambers
Interfaces Ethernet or other common fi eldbus systems and serial interfaces
Dimensions 1200 x 975 x 2260 mm (W x D x H), Widths 900 mm, 1500 mm and 1800 mm also available
Weight (without equipment) 260 kg for 1200 mm width (320 kg, 380 kg for wider models)
Dimensions working area 1180 x 690 x 880 mm (W x D x H), width varies with total width
Working height 700 mm, 900 mm
Front sash opening height Individually adjustable via PLC

Contact

Contact us for guidance on your project

Talk to our experts.

More products

in the category of metal-free worstations