Combination Station vertical

System details
- Automatic vertical front sash (clear)
- Light curtain for activity monitoring
- Integrated LED lights
- Monitoring of differential pressure and filter loading
- Evaporation chamber with integrated pass-through
- Temperature monitoring of exhaust air
- Monitoring of temperature
- Programmable logic controller (PLC) for monitoring and controlling of operation, airflows and equipment
- Supply air filters according to ISO class
- Volume flow controllers for supply and exhaust air
- Connections for hotplates
- Bottling station (various versions)
- Exhaust tubflushing
- Drain / wastewater connection
- Customized media supply (interior and exterior)
- Remote monitoring and servicing
- Integrated ultrapure water system (various manufacturers)
- Integrated acid stills
- Ultrasonic basin
- Base cabinet (vented) with hinged doors or drawers
- Exhaust air scrubber and exhaust air irrigation

Laminar flow
DIN EN 14175
Modular design
Reduce energy consumption
MK Versuchsanlagen combination Station vertical
Our metal-free systems feature a modular design, allowing us to create custom-made solutions tailored to your specific needs:
-
Top module: Controlled ventilation system with filtered laminar airflow
-
Middle module: Customizable work surface and front sash
-
Base module: Flexible substructure for additional storage space or legroom
The metal-free construction enables contamination-free sample preparation at the highest level, ideal for trace analysis, for example. The station consists of two self-contained work areas with laminar airflow: the work chamber and the evaporation chamber, which are connected by a pass-through. The two autonomous chambers allow the evaporation processes to be separated from subsequent steps, ensuring improved personnel protection.
Additional features
-
Metal-free construction (no exposed metal) to prevent contamination and corrosion
-
Supply air via available supply air from in-house ventilation system
-
Vertical laminar flow achieved though a fine-meshed monofilament fabric diffuser
-
Laminar airflow throughout the entire working chamber; air is exhausted through perforated work surface
-
Control of supply air and exhaust air via the PLC ensures the correct pressure in the working chamber
-
Exhaust air is extracted through a perforated fl oor via underfloor basin
- Compliant with DIN EN 14175-3 safety and performance standards (containment, air exchange effi ciency)
- Simultaneous protection of personnel and samples/products
- Outstanding air and energy effi ciency due to optimized and smart airfl ow management
- Exhaust air can be discharged via air scrubbers or directly transferred to the exhaust air system
- Room air can also be extracted via the base module and fed into the ventilation system (‘recirculating air’)
- Separate chamber (with independent vertical laminar flow) for isolated evaporation / hotplate usage
Technical data | ||
Materials | PP, PET, POM, PFA, PVDF | |
Operating voltage | 400 V, 5-pole | |
Air volume requirements |
Supply air min. 400 m³/h, exhaust air min. 450 m³/h (Width: 1200 mm) |
|
Flow velocity in working area | approx. 0.15 m/s | |
Terminal supply air filters | High-performance HEPA/ULPA filter, class H13 to U16 (depending on requirements) | |
Cleanroom class according DIN EN ISO 14644-1 |
Class 3 or better (0.2 μm, 0.3 μm, 0.5 μm, 1 μm, 5 μm) „at rest“ | |
Illuminance | > 750 Lux in both chambers | |
Interfaces | Ethernet or other common fi eldbus systems and serial interfaces | |
Dimensions | 1200 x 975 x 2260 mm (W x D x H), Widths 900 mm, 1500 mm and 1800 mm also available | |
Weight (without equipment) | 260 kg for 1200 mm width (320 kg, 380 kg for wider models) | |
Dimensions working area | 1180 x 690 x 880 mm (W x D x H), width varies with total width | |
Working height | 700 mm, 900 mm | |
Front sash opening height | Individually adjustable via PLC |
Contact